fbpx

JL3KH6 – Job Link® System Charge and Air Kit

Quick service checks just got quicker.

Wireless range of 1,000’.

Where to Buy

Directly connects to mobile device with Job Link System App
Fast accurate temps with new Rapid Rail technology
Pressures auto calibrate at any elevation
Configurable psychrometers with flexible probe and sliding magnet
A2L Compatible


Take a full system snapshot quickly and easily with the JL3KH6 Charge and Air Test Kit. This kit consists of 2 pressure probes, 2 pipe clamps, and 2 psychrometers in a quick access padded case. All measurements can be sent up to 1000’ away directly to your mobile device with the Job Link® System mobile app. Each tool is designed for agility and speed. The pressure probes fit into tight spaces. The pipe clamps use the new Rapid Rail™ sensor design delivering accuracy, speed, and agility, even on the most awkward pipe configurations. The psychrometers’ flexible narrow probes and configurable magnet make it easy to measure supply plenums, ducts, and elevated registers. Do more with Job Link probes!

SPECIFICATIONS

Wireless Range:    1000 feet or more
Wireless Frequency:    2.4 GHz
Battery:    2 x AAA (included)
Battery Life:    150 hrs typical alkaline
Auto Power Off (APO):    2 hrs (can be disabled)
Minimum Device Requirement:    BLE 4.0 devices running iOS® 7.0 or AndroidTM 5.0

Job Link Pressure Probe – JL3PR
Connector Type:    45° Standard 1/4” NPT female flare
Best Accuracy:    ±1 psig (±7 kPa)
Measurement Range:    0 to 580 psig (0 to 4000 kPa)
Maximum Overload:    800 psig (5500 kPa)
Water Resistant:    Designed to IP55

Job Link Premium Pipe Clamp Probe – JL3PC
Pipe Compatibility:    Electrically conductive 1/4” to 1 3/8” (6.4mm to 34.9mm) dia.
Accuracy:    ±1°F (±0.6°C)
Measurement Range:    -50°F to 257°F (-46°C to 125°C)
Stabilization Time:    3 seconds typical
Measurement Feedback:    Beeper and LED
Water Resistant:    Designed to IP55
Patent:    Pending

Job Link Flex Psychrometer Probe – JL3RH
Flexible Probe:    0.36” (9mm) diameter, 9.25” (235mm) length
Best Accuracy:    ±1°F (±0.6°C) ±2.5 %RH
Measurement Range:    -40°F to 250°F (-40°C to 121°C) 0 %RH to 100 %RH

Includes:

  • 2 Job Link Pipe Temp Probes
  • 2 Job Link Pressure Probes
  • 2 Job Link Psychrometer Probes
  • Padded Carrying Case
  • 2 x AAA Batteries Per Probe (12 Total)

Read More

Specifications User Manuals Print Material FAQs

See the Power of Job Link First Hand

Where to Buy

JL3KH6 – Job Link® System Charge and Air Kit

Quick service checks just got quicker.

Wireless range of 1,000’.

Directly connects to mobile device with Job Link System App
Fast accurate temps with new Rapid Rail technology
Pressures auto calibrate at any elevation
Configurable psychrometers with flexible probe and sliding magnet
A2L Compatible


Take a full system snapshot quickly and easily with the JL3KH6 Charge and Air Test Kit. This kit consists of 2 pressure probes, 2 pipe clamps, and 2 psychrometers in a quick access padded case. All measurements can be sent up to 1000’ away directly to your mobile device with the Job Link® System mobile app. Each tool is designed for agility and speed. The pressure probes fit into tight spaces. The pipe clamps use the new Rapid Rail™ sensor design delivering accuracy, speed, and agility, even on the most awkward pipe configurations. The psychrometers’ flexible narrow probes and configurable magnet make it easy to measure supply plenums, ducts, and elevated registers. Do more with Job Link probes!

SPECIFICATIONS

Wireless Range:    1000 feet or more
Wireless Frequency:    2.4 GHz
Battery:    2 x AAA (included)
Battery Life:    150 hrs typical alkaline
Auto Power Off (APO):    2 hrs (can be disabled)
Minimum Device Requirement:    BLE 4.0 devices running iOS® 7.0 or AndroidTM 5.0

Job Link Pressure Probe – JL3PR
Connector Type:    45° Standard 1/4” NPT female flare
Best Accuracy:    ±1 psig (±7 kPa)
Measurement Range:    0 to 580 psig (0 to 4000 kPa)
Maximum Overload:    800 psig (5500 kPa)
Water Resistant:    Designed to IP55

Job Link Premium Pipe Clamp Probe – JL3PC
Pipe Compatibility:    Electrically conductive 1/4” to 1 3/8” (6.4mm to 34.9mm) dia.
Accuracy:    ±1°F (±0.6°C)
Measurement Range:    -50°F to 257°F (-46°C to 125°C)
Stabilization Time:    3 seconds typical
Measurement Feedback:    Beeper and LED
Water Resistant:    Designed to IP55
Patent:    Pending

Job Link Flex Psychrometer Probe – JL3RH
Flexible Probe:    0.36” (9mm) diameter, 9.25” (235mm) length
Best Accuracy:    ±1°F (±0.6°C) ±2.5 %RH
Measurement Range:    -40°F to 250°F (-40°C to 121°C) 0 %RH to 100 %RH

Includes:

  • 2 Job Link Pipe Temp Probes
  • 2 Job Link Pressure Probes
  • 2 Job Link Psychrometer Probes
  • Padded Carrying Case
  • 2 x AAA Batteries Per Probe (12 Total)

SPECIFICATIONS

  • Wireless Range: 1000 feet or more (305 meters)
  • Wireless Frequency: 2.4 GHz
  • Battery: 2x AAA (included)
  • Battery Life: 150 hours typical alkaline
  • Auto Power Off (APO): 2 hours (can be disabled)
  • Minimum Device Requirement: BLE 4.0 devices running iOS® 7.0 or Android™ 5.0

Job Link Pressure Probe – JL3PR

  • Connector Type: 45° Standard 1/4” NPT female flare
  • Best Accuracy: ±1 psig (±7 kPa)
  • Measurement Range: 0 to 580 psig (0 to 4000 kPa)
  • Maximum Overload: 800 psig (5500 kPa)
  • Water Resistant: Designed to IP55

Job Link Premium Pipe Clamp Probe – JL3PC

  • Pipe Compatibility: Electrically conductive 1/4” to 1 3/8” (6.4mm to 34.9mm) dia.
  • Accuracy: ±1°F (±0.6°C)
  • Measurement Range: -50°F to 257°F (-46°C to 125°C)
  • Stabilization Time: Three seconds typical
  • Measurement Feedback: Beeper and LED
  • Water Resistant: Designed to IP55
  • Patent: 10,365,165

Job Link Flex Psychrometer Probe – JL3RH

  • Flexible Probe: 0.36” (9mm) diameter, 9.25” (235mm) length
  • Best Accuracy: ±1°F (±0.6°C) ±2.5 %RH
  • Measurement Range: -40°F to 250°F (-40°C to 121°C) 0 %RH to 100 %RH

Opman JL3PR v17 Opman JL3PR v17

Opman JL3PC v17 Opman JL3PC v17

Opman JL3RH v17 Opman JL3RH v17

  • What is the range of JL3PR?

    • 1000 FT line of sight

  • Why am I not getting a superheat or subcooling measurement

    • There are three possibilities.

      • No pressure probe connected
      • No pipe clamp connected
      • No refrigerant selected
  • I have a 2-stage system, can I see both compressors at the same time?

    • Yes, hook up the probes you will be using and push the blue serial number next to the probe you would like to switch and select the probe you would like to see.

  • What is the range of JL3PC?

    • 1000 FT line of sight

  • Why does it show open when pipe clamp is closed

    • The pipe that the clamp is on may to need to be cleaned. Remove the clamp and use the Emory cloth provided to clean the pipe and put clamp back on.

  • Why are my temperature extreme (1000+F)

    • The probes update at a rapid speed between opening/closing, it uses an extreme value to deter the possibility of catching a miscalculation. It should go back to its actual temperature within seconds

  • What is the range of JL3RH?

    • 1000 FT line of sight

  • Is the JL3KH6 A2L compatible?

    • Yes. For use only by qualified and certified technicians in the safe use, handling, and transporting of refrigerants.  Please refer to flammable refrigerant safety guides, regional codes and legislation for more information.